Setting Up Marco ARK, HandyRef, LM, RKT, & VL Series Equipment

The following models interact with ExamWRITER without a digital refraction system:

  • ARK-1
  • ARK-1a
  • ARK-1e
  • ARK-530A*
  • ARK-700A*
  • ARK-710A*
  • ARK-730A*
  • ARK-760A*
  • ARK-F*
  • LM-7**
  • LM-7P**
  • LM-500**
  • LM-600**
  • LM-990a**
  • LM-1000**
  • LM-1200**
  • LM-1800**
  • RKT-7700**
  • M3 TonoRef II
  • VL-3000

* Select ARK-A-Devices from the Marco device list in the Eyefinity Equipment Interface Setup window.

** Select LM-Devices from the Marco device list in the Eyefinity Equipment Interface Setup window.

 

Daisy-chaining (connecting lensometers to autorefractors) is not supported.

For the device to transfer data to the equipment interface, you must set several parameters on your Marco equipment as follows:

  1. To access the Parameter Set screen and set each of the parameters, follow the instructions in your device manual.
  2. When you finish setting the parameters, exit Parameter Set mode, turn off the device, and turn it back on.

The following table shows the parameter number and setting for each of the parameters on each device.

Device KM Unit I/F Mode I/F Format Baud Rate Bit Length CR Code LDATA

ARK-1

#3: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

No

ARK-1a

#3: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

No

ARK-1e

#3: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

No

ARK-530A

 

#71: NCP10

#72: All

 

 

 

 

ARK-700A

#3: D

#31: NCP10

#32: All

#33: 9600

#34: 8

35: No

 

ARK-710A

#11: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

 

ARK-730A

#11: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

 

ARK-760A

#3: D

#51: NCP10

#52: All

#53: 9600

#54: 8

55: No

 

ARK-F

To set these parameters on this device, use the Marco Connect software. To download the Marco Connect software, contact Marco at 800.874.5274 or go to www.marcoconnect.com/downloads.

HandyRef-K

 

#71: NCP10

#72: All

 

 

 

 

LM-7

 

NCP10

 

 

 

 

 

LM-7P

 

NCP10

 

 

 

 

 

LM-600

 

Menu Page 5: #32 NCP10

 

#33: 9600

8

OFF

 

LM-990A

To set these parameters on this device, use the Marco Connect software. To download the Marco Connect software, contact Marco at 800.874.5274 or go to www.marcoconnect.com/downloads.

LM-1000

 

Menu Page 5: NCP10

 

Menu Page 5: 9600

Menu Page 5: 8

Menu Page 5: No

 

LM-1200

 

Menu Page 4: NCP10

 

Menu Page 4: 9600

Menu Page 4: 8

Menu Page 4: No

 

LM-1800

 

Menu Page 5: #32 NCP10

 

#33: 9600

8

OFF

 

RKT-7700

 

#49: NCP10

#50: All

 

 

 

 

M3 TonoRef II

 

#81: NCP10

#82: All

 

 

 

 

VL-3000

PC

9600

8

ODD

1

VL-3000