Setting Up the Reichert Interface
Before setting up the Reichert interface, note the following:
- You must set up the interface from the computer connected to the Reichert equipment.
- If you use the AT-555 Tonometer, make sure the device’s firmware is version 2.4 or later. For information on checking the firmware version, see the documentation that came with the device.
- Make sure you have the proper serial cable for your equipment:
For this device . . . | Use this cable . . . |
---|---|
AL-200 Lensometer | NULL modem |
RK-600 Autorefractor | Reichert custom cable |
iPac Pachymeter | Wireless |
All other Reichert devices | Standard serial cable |
To set up the Reichert interface
- Do one of the following:
- If you are not setting up the iPac interface, see the documentation that came with your device to set the RS-232 data bus parameters listed in the following table.
- If you are setting up the iPac interface, go to step 3.
- See the documentation that came with your device to set the baud rate:
Model Baud Rate AL-200, PT-100, RK-600
9600
AL-500
38400
AL-700
38400 (fixed at 921600 when USB is selected)
AT-555, Reichert 7, Reichert 7CR
19200
RK-700
115200 (fixed at 921600 when USB is selected)
- In the ExamWRITER main window, click Tools, select Equipment Integration Setup, and select Equipment Interface Setup.
The Eyefinity Equipment Interface Setup window opens.
- Click Add.
- Select the Reichert manufacturer and a Reichert interface device:
For this model . . . Select a Reichert interface device . . . AL-200
AL-200 (Lensmeter) AL-500, AL-700, LensChek, or OptoChek Lensometer (Lensmeter) AT-555
AT-555 (Tonometer) PT-100, Reichert 7, or Reichert 7CR Tonometer (Tonometer) RK-600 RK-600 (Auto Refractor} RK-700 Auto Refractor (Auto Refractor) Auto Phoroptor RS, Phoroptor VRx, SightChek, or Visutron900 not linked to an autorefractor or lensmeter Auto Phoropter Output (Auto Phoropter) and
Auto Phoropter Input (Auto Phoropter)Auto Phoroptor RS, Phoroptor VRx, SightChek, or Visutron900 linked to an autorefractor or lensmeter Auto Phoroptor Output (Auto Phoropter) SightChek Input
SightChek Input (Auto Phoropter)
To enable ExamWRITER to send data in XML format to the Reichert SightChek forepart, select the following options in the Eyefinity Equipment Interface Setup window:
Manufacturer: Reichert
Device: SightChek Input (Auto Phoropter)
Export Folder: The shared folder set up in the ReichertCapture software for the SightCheck forepart.
Click Save, and then close the Equipment Interface Setup window.
iPac Pachymeter iPac (Pachymeter) Reflex UBM UBM Image Review (Ultrasound Biomicroscope) Foresee PHP PHP Image Review (Perimeter) - Click Add.
The settings and options appear.
- (Optional) Enter a Device Nickname.
- If you are setting up an AL-200, AT-555, or RK-600 device, do one of the following:
To connect to the computer by a serial cable:
Select the Use Serial Port option.
From the Select COM Port drop-down list, select a port number.
From the Select Baud Rate drop-down list, select the baud rate.
To receive input from Reichert Capture:
Select the Reichert Capture option.
Click Browse, and select the drive and folder (Reichert Capture Folder) where the equipment data is temporarily stored.
For all other Reichert devices except input autophoroptors, click Browse, and select the location of the exchange folder in which the temporary data file is saved.
You must select an empty folder that is not used by any other application, including Reichert. If you select a folder that contains files, they will be deleted during the integration and data transfer process.
In the File Prefix field, enter the beginning characters of the data file as shown in the following table.
Reichert Model Data File Start Chars AL-200
AL-200
AL-500
AL-500
AL-700
AL-700
AT-555
AT-555
iPac Pachymeter
iPac
LensChek Plus Digital Lensometer
Reichert
LensChek Pro Digital Lensometer
lenscheckpro
OptoChek Plus Autorefractor/Keratometer
Reichert
Reichert7
Reichert7
Reichert7CR
Reichert7CR
RK-600
RK-600
RK-700
RK-700
Auto Phoroptor RS
Auto
Phoroptor VRx
Phoroptor
SightChek
sightchek
Visutron 900
Auto
- From the File Extension drop-down list, select one of the following:
- For a UBM Image Review or PHP Image Review device, select .jpg.
- For any other equipment interface except the Auto Phoroptor Input, select .xml.
- For the Auto Phoroptor Input, do not enter an extension.
- Click Save.
- Close the Eyefinity Equipment Interface Setup window.
Model | Parameter | Settings | Notes |
---|---|---|---|
All | Character | 8 | Required |
All | Parity | N | Required |
All | Stop bits | 1 | Required |
AL-200 | PRSM | Base | Required to record the presenting horizontal and vertical prism data |
RK-600 | KERATO | Diopt | Required to record the kerotometry readings in ExamWRITER |
RK-600 | Manual | Required | |
RL-600 | REVIEW DATA | On | Required |